135
3.3.2.6. UE Transmit OFF power for CA (intra-band contiguous DL CA and UL CA) (6.3.3A.1)
Refer to Chapter 3.3.2.7.
3.3.2.7. General ON/OFF time mask for CA (intra-band contiguous DL CA and UL CA) (6.3.4A.1.1)
3.3.2.7.1. MT8820C
This measurement item is not supported by the MT8820C.
Refere to Table 3.3–1.
3.3.2.7.2. MT8821C
This subsection describes an example of intra–band measurement.
[Pass/Fail evaluation limits value setting]
1. Execute TP_OFFPWR_UL -48.5, PCC to set TX2 – General Time Maskof Off Power Pass/Fail judgment for PCC.
2. Execute TP_TMASK_GEN_TOL 7.5, PCC to set TX2 – General Time Maskof On Power Pass/Fail judgment for PCC.
3. Execute TP_OFFPWR_UL -48.5, SCC1 to set TX2 – General Time Maskof Off Power Pass/Fail judgment for SCC1.
4. Execute TP_TMASK_GEN_TOL 7.5, SCC1 to set TX2 – General Time Maskof On Power Pass/Fail judgment for SCC1.
[Measurements]
5. Execute TESTPRM TX_GEN_TMASK to set Test Parameter to TX2 – General Time Mask.
6. Execute PT_WDR ON to enable Power Template Wide Dynamic Range measurement.
7. Execute SWP to measure the Power Template.
The following procedure is for validating the PCC result.
8. Execute ONPWR? AVG, PCC to read the On Power measurement result for PCC.
9. Execute ONPWRPASS? PCC to check that the On Power measurement Pass/Fail judgment for PCC is Pass.
10. Execute OFFPWR_BEFORE? AVG, PCC to read the Off Power (Before) measurement result for PCC.
11. Execute OFFPWR_AFTER? AVG, PCC to read the Off Power (After) measurement result for PCC.
12. Execute OFFPWRPASS? PCC to check that the Off Power measurement Pass/Fail judgment for PCC is Pass.
The following procedure is for validating the SCC–1 result.
13. Execute ONPWR? AVG, SCC1 to read the On Power measurement result for SCC–1.
14. Execute ONPWRPASS? SCC1 to check that the On Power measurement Pass/Fail judgment for SCC–1 is Pass.
15. Execute OFFPWR_BEFORE? AVG, SCC1 to read the Off Power (Before) measurement result for SCC–1.
16. Execute OFFPWR_AFTER? AVG, SCC1 to read the Off Power (After) measurement result for SCC–1.
17. Execute OFFPWRPASS? SCC1 to check that the Off Power measurement Pass/Fail judgment for SCC–1 is
Pass.
NOTE: Pass/Fail judgment values for Transmitted Off Power differ depending on the Carrier Frequency f.
f ≤ 3.0 GHz: ≤–48.5 dBm (initial value)
3.0GHz < f ≤ 4.2 GHz: ≤–48.2 dBm