3.3.2.10. Aggregate power control tolerance for CA (intra-band contiguous DL CA and UL CA)
(6.3.5A.3.1)
3.3.2.10.1. MT8820C
This measurement item is not supported by the MT8820C.
Refere to Table 3.3-1.
3.3.2.10.2. MT8821C
This subsection describes an example of intra–band measurement.
[Measurements]
1. Execute TESTPRM TX_PCTAGG_PUSCH to set Test Parameter to TX3 – Aggregate Power (PUSCH Sub–test).
2. Execute SWP to measure Power Control Tolerance (Aggregate Power).
3. Execute PCTPWR? PCC to read the Aggregate Power (dB) measurement result for PCC.
4. Execute PCTPASS? PCC to check that the Aggregate Power measurement Pass/Fail judgment for PCC is Pass.
5. Execute PCTPWR? SCC1 to read the Aggregate Power (dB) measurement result for SCC–1.
6. Execute PCTPASS? SCC1 to check that the Aggregate Power measurement Pass/Fail judgment for SCC–1 is
Pass.
7. Execute PCTPASS? to check that the Aggregate Power measurement Pass/Fail judgment for all CCs is Pass.
8. Execute TESTPRM TX_PCTAGG_PUCCH to set Test Parameter to TX3 – Aggregate Power (PUCCH Sub–
test).
9. Execute SWP to measure the Power Control Tolerance (Aggregate Power).
10. Execute PCTPWR? to read the Aggregate Power (dB) measurement result.
11. Execute PCTPASS? to check that the Aggregate Power measurement Pass/Fail judgment is Pass.