137
3.3.2.9. Power Control Relative power tolerance for CA (intra-band contiguous DL CA and UL CA)
(6.3.5A.2.1)
3.3.2.9.1. MT8820C
This measurement item is not supported by the MT8820C.
Refere to Table 3.3-1.
3.3.2.9.2. MT8821C
This subsection describes an example of intraband measurement.
[Pass/Fail evaluation limits value setting]
1. Execute TP_PCTREL_RMP_TOL 1.7, PCC to set TX3 Relative Power (Ramping Up/Down) Pass/Fail tolerance
of subframes before/after RB change for PCC.
2. Execute TP_PCTREL_RMP_TOL 1.7, SCC1 to set TX3 Relative Power (Ramping Up/Down) Pass/Fail
tolerance of subframes before/after RB change for SCC1.
3. Execute TP_PCTREL_RMP_CNG_TOL1 4.7, PCC to set TX3 Relative Power (Ramping Up/Down) RB Change
Pass/Fail tolerance for PCC.
4. Execute TP_PCTREL_RMP_CNG_TOL1 4.7, SCC1 to set TX3 Relative Power (Ramping Up/Down) RB Change
Pass/Fail tolerance for SCC-1.
5. Execute TP_PCTREL_RMP_CNG_TOL2 5.7, PCC to set TX3 Relative Power (Ramping Up/Down) RB Change
Pass/Fail tolerance for PCC.
6. Execute TP_PCTREL_RMP_CNG_TOL2 5.7, SCC1 to set TX3 Relative Power (Ramping Up/Down) RB Change
Pass/Fail tolerance for SCC-1.
7. Execute TP_PCTREL_RMP_CNG_TOL3 6.7, PCC to set TX3 Relative Power (Ramping Up/Down/Down) RB
Change Pass/Fail tolerance for PCC.
8. Execute TP_PCTREL_RMP_CNG_TOL3 6.7, SCC1 to set TX3 Relative Power (Ramping Up/Down/Down) RB
Change Pass/Fail tolerance for SCC-1.
9. Execute TP_PCTREL_RMP_E 6.7, PCC to set TX3 Relative Power (Ramping Up/Down) tolerance of the
Exception points for PCC.
10. Execute TP_PCTREL_RMP_E 6.7, SCC1 to set TX3 Relative Power (Ramping Up/Down) tolerance of the
Exception points for SCC-1.
11. Execute TP_PCTREL_ALT_TOL 6.7, PCC to set TX3 Relative Power (Alternating) Pass/Fail tolerance for PCC.
12. Execute TP_PCTREL_ALT_TOL 6.7, SCC1 to set TX3 Relative Power (Alternating) Pass/Fail tolerance for
SCC1.
[Measurements]
1. Execute TESTPRM TX_PCTREL_UP_A to set Test Parameter toTX3 Relative Power (Ramping Up A).
2. Execute SWP to measure Power Control Tolerance (Relative Power).
3. Execute PCTPWR? PCC to read the Relative Power (dB) measurement result for PCC.
4. Execute PCTPWR2? PCC to read the Relative power (dB) measurement result at RB Change for PCC .
5. Execute PCTPASS? PCC to check that the Relative Power measurement Pass/Fail judgment for PCC is Pass.
6. Execute PCTPWR? SCC1 to read the Relative Power (dB) measurement result for SCC1.
7. Execute PCTPWR2? SCC1 to read the Relative power (dB) measurement result at RB Change for SCC1 .
8. Execute PCTPASS? SCC1 to check that the Relative Power measurement Pass/Fail judgment for SCC1 is
Pass.
9. Execute PCTPASS? to check that the Relative Power measurement Pass/Fail judgment for all CCs is Pass.
10. Execute TESTPRM TX_PCTREL_UP_B to set Test Parameter to TX3 Relative Power (Ramping Up B).
11. Execute steps 2 to 9.
12. Execute TESTPRM TX_PCTREL_UP_C to set Test Parameter to TX3 Relative Power (Ramping Up C).
13. Execute steps 2 to 9.
14. Execute TESTPRM TX_PCTREL_DOWN_A to set Test Parameter to TX3 Relative Power (Ramping Down
A).
15. Execute steps 2 to 9.
16. Execute TESTPRM TX_PCTREL_DOWN_B to set Test Parameter toTX3 Relative Power (Ramping Down B).
17. Execute steps 2 to 9.
18. Execute TESTPRM TX_PCTREL_DOWN_C to set Test Parameter to TX3 Relative Power (Ramping Down
C).
19. Execute steps 2 to 9.
20. Execute TESTPRM TX_PCTREL_ALT to set Test Parameter toTX3 Relative Power (Alternating).
21. Execute steps 2 to 9 except 4 and 7.
NOTE 1: The tolerance of RB Change subframe is varied for transmission bandwidths confined within
F
UL_low
and F
UL_low
+ 4 MHz or F
UL_high
4 MHz and F
UL_high
.