144
3.3.2.15. Occupied bandwidth for CA (intra-band contiguous DL CA and UL CA) (6.6.1A.1)
3.3.2.15.1. MT8820C
This measurement item is not supported by the MT8820C.
Refer to Table 3.3–1.
3.3.2.15.2. MT8821C
This subsection describes an example of intra–band measurement.
[Measurements]
1. Execute OBW_AVG 20 to set the average count for Occupied Bandwidth to 20 times.
2. Execute TESTPRM TX_MAXPWR_Q_F to set Test Parameter to TX1 – Max. Power (QPSK/FullRB).
3. Execute SWP to measure the Occupied Bandwidth (OBW).
4. Execute OBW? to read the OBW measurement result.
5. Execute OBWPASS? to check that the OBW Pass/Fail judgment is Pass.