AX-4030
4-13
AUTION MAX AX-4030 OPERATING MANUAL
D
3 Tilt the feeder unit down to the right.
1 Tilt the feeder unit to the right.
The feeder unit is locked with the latch shown with
the circles in the figure. So, first you need to pull
the whole feeder unit to the right with a little force
to open the latch.
NOTE: Do not apply excessive pressure to the
feeder unit when in the tilted position, as this may
cause damage to the instrument.
4 Remove the introduction tray.
1 Pull out the introduction tray about 20 mm in the
direction of the arrow along the introduction arms.
2 Taking care not to hit the tray against the arms, lift
the introduction tray up at an angle and remove the
tray.
NOTE: Avoid strong impact to the introduction arms
while removing the introduction tray. These arms
are made of fragile materials.
NOTE: Be careful not to touch the nozzle while
removing the introduction tray.
5 Wash and sterilize the introduction tray.
1 Sterilize the introduction tray with alcohol.
2 Rinse the tray using a sponge to thoroughly remove
dirt.
NOTE: Do not use hot water to rinse the tray, as it
may deform the tray.
NOTE: Protect the tray from scratches, which may
prevent smooth transportation of test strips.
NOTE: Some portions of the tray attract more dirt
than others. Rinse the tray thoroughly until the
whole tray is visually clean.
3 Use a cloth to wipe water off the tray, and allow it to dry.
Feeder
unit
Introduction arms
Introduction tray