AX-4030
4-13
AUTION MAX AX-4030 OPERATING MANUAL
D
3 Tilt the feeder unit down to the right.
1 Tilt the feeder unit to the right.
• The feeder unit is locked with the latch shown with
the circles in the figure. So, first you need to pull
the whole feeder unit to the right with a little force
to open the latch.
NOTE: Do not apply excessive pressure to the
feeder unit when in the tilted position, as this may
cause damage to the instrument.
4 Remove the introduction tray.
1 Pull out the introduction tray about 20 mm in the
direction of the arrow along the introduction arms.
2 Taking care not to hit the tray against the arms, lift
the introduction tray up at an angle and remove the
tray.
NOTE: Avoid strong impact to the introduction arms
while removing the introduction tray. These arms
are made of fragile materials.
NOTE: Be careful not to touch the nozzle while
removing the introduction tray.
5 Wash and sterilize the introduction tray.
1 Sterilize the introduction tray with alcohol.
2 Rinse the tray using a sponge to thoroughly remove
dirt.
NOTE: Do not use hot water to rinse the tray, as it
may deform the tray.
NOTE: Protect the tray from scratches, which may
prevent smooth transportation of test strips.
NOTE: Some portions of the tray attract more dirt
than others. Rinse the tray thoroughly until the
whole tray is visually clean.
3 Use a cloth to wipe water off the tray, and allow it to dry.