4-28
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
D
4 Clean the washing bath.
1 Wipe dirt off inside the washing bath and port using
cotton swabs moistened with purified water.
• Pay additional attention when cleaning the two level
detection sensors in the washing port.
NOTE: Do not touch the nozzle while cleaning the
washing bath and port.
NOTE: After cleaning, check inside the washing
bath and port for lint or other dust, which may
cause the flow line filter to clog.
5 Close the cover and turn on the power.
1 Close the front cover.
2 Press the standby switch to turn on the power.
• Allow the instrument to warm up.
• The nozzle will automatically return to its home
position.