EasyManuals Logo
Home>Arkray>Measuring Instruments>AUTION MAX AX-4030

Arkray AUTION MAX AX-4030 User Manual

Arkray AUTION MAX AX-4030
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #171 background imageLoading...
Page #171 background image
4-28
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
D
4 Clean the washing bath.
1 Wipe dirt off inside the washing bath and port using
cotton swabs moistened with purified water.
Pay additional attention when cleaning the two level
detection sensors in the washing port.
NOTE: Do not touch the nozzle while cleaning the
washing bath and port.
NOTE: After cleaning, check inside the washing
bath and port for lint or other dust, which may
cause the flow line filter to clog.
5 Close the cover and turn on the power.
1 Close the front cover.
2 Press the standby switch to turn on the power.
Allow the instrument to warm up.
The nozzle will automatically return to its home
position.
Washing port
Level detection sensors
(Clean with great care.)
Washing bath
Standby
switch
Front cover