AX-4030
5-5
AUTION MAX AX-4030 OPERATING MANUAL
W010 Measurement cannot start. Perform S.G. calibration to enable measurement.
Cause The existing S.G. calibration result was invalidated after the specified period of time elapsed
since the last calibration.
Remedy
Press OK to clear the warning.
Perform S.G. calibration to update the result (see page 2-34), and retry the tests.
W011 The test strips are not sent out of the feeder correctly.
To continue measurement, mix the test strips in the feeder.
Cause Several test strips are trapped inside the feeder or sticking together.
Remedy Mix the test strips in the feeder.
Press
OK to clear the warning.