AUTION MAX AX-4030 OPERATING MANUAL
E130 Two-way communication error
Cause The communication cable is not connected correctly to the external device.
Remedy
Press OK to clear the error.
Securely connect the communication cable to both the instrument and host.
E140 Test strip marker error
Cause The test strip type detected from the strip marker and the test strip type setting do not match.
Remedy
Press OK to clear the error.
Make the correct setting for the type of test strips loaded into the feeder (see page 2-16).
Check the feeder contains the correct type of test strips, and remove the wrong types of test
strips, if there are any.