1-24
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
6 Feed the sampler cables through the left side of the
instrument.
7 Loosen the temporarily-tightened screw using a
Phillips screwdriver, and slide the splash guard
down.
NOTE: Make sure the cables do not get caught in
the splash guard.
8 Tighten the two screws using a Phillips screwdriver to secure the splash guard.
3 Attach the sampler to the instrument.
1 Fit the hooks on both sides of the sampler into the
holes on the instrument.
Be careful not to pinch the cables and earth line.
2 Loosen the locking nut on the instrument legs close
to the sampler.
3 Turn the adjuster feet until the sampler rises an even
5 mm off the countertop.
Check that the instrument has a clearance of about
20 mm from the countertop.
4 Tighten the locking nuts to lock the adjuster feet in place.
Screw holes
Splash guard
Cables
Hook
Instrument Sampler
5 mm
About 20 mm
Turn here to adjust height.
Loosen
this nut.