AX-4030
2-17
AUTION MAX AX-4030 OPERATING MANUAL
D
3 Below [Test strips in feeder 1] and [Test strips in
feeder 2], select the type of test strips to be loaded
into each feeder.
• Press the hyphen key to change the options.
• Press to move the cursor down.
4 Below [Main feeder], select the feeder you want to
use as the main feeder.
• Press the hyphen key to switch between feeders 1
and 2.
5 Press OK .
• A message window will open to ask if you want to
save your settings.
6 Press OK to save your changes.
• The standby screen will appear with the new
settings displayed.
REFERENCE: Press Cancel to abort your changes
and return to the standby screen.
IMPORTANT: Keep the instrument on while
saving the settings.
2 Attach a desiccant bag to the feeder cover.
1 Turn the locking lever (a) to unlock the feeder
cover, and open the cover (b).