2 Tilt the side panel of the storage section in the
direction of the arrow.
3 Take out a desiccant bag from the test strip bottle
and attach it to the feeder cover.
• For loading 100 test strips, attach a desiccant bag on
the side panel of the storage section. If loading 200
test strips, attach one bag on the back of the cover,
and another bag on the side panel of the storage
section.
4 Place the side panel as before.
REFERENCE: Replace the desiccant bags on
the feeders each time a new test strip bottle is
opened.
3 Load the test strips into the feeders.
1 Take out as many test strips as you need from the
bottle, and load them into the feeder.
• Level test strips into even piles.
IMPORTANT: Do not touch the pad area on
the test strips. Contact can cause sebum to
adhere to the surface, whereby producing
incorrect results.
NOTE: Place the test strips with the black marker to
the back side of the feeder (see the figure at right).
2 Close the feeder cover and turn the locking lever clockwise to lock it.
3 Cap the test strip bottle.
IMPORTANT: Without desiccant bags, the test
strips can absorb moisture in the air. Cap the
bottle securely to avoid deterioration of the
strips.