Calibration
119
1(1)
TRANSDUCER CALIBRATION REPORT
Transducer serial number Date Time
20529 March 28, 1999 10:55.12
ACTA SERIAL NUMBER 1234567890
OTHER EQUIPMENT
PREVIOUS CALIBRATION DATE Sept 08, 1998
TRANSDUCER TYPE IRTT WITH ANGLE ENCODER
SENSITIVITY 2.0 mV/V
CALIBRATION INPUT No Load Cal point Cal value
Nm) (mV/V) (Nm) (mV/V) (Nm)
0 -0.05 175 1.98 172.4
0 -0.02 175 1.95 177.7
0 -0.05 175 1.91 178.6
STANDARD DEVIATION 2.74
LINEARITY CHECK True value Measured value Unlinear
(Nm) (Nm) (%)
87 89.8 +1.59
132 130.3 -0.97
STANDARD DEVIATION 1.32
OLD CALIBRATION VALUE 173.8 Nm
NEW CALIBRATION VALUE 176.1 Nm SAVED IN TRANSDUCER .... ....
YES NO
NEXT CALIBRATION DATE ............................................
ADDITIONAL INFORMATION ............................................
............................................
............................................
SIGNATURE
....................