Programming
52
§ Calibration date
§ Calibration source (ACTA 3000 or other calibration equipment)
§ Next calibration date
§ Last update of calibration torque (date)
§ Signature of calibrator
§ Comments
Note!
When an Atlas Copco transducer with memory is connected to the ACTA 3000, a check of the
next calibration date is performed:
§ If there are less than 30 days to the next calibration date, a dialog box appears telling the operator
that the transducer needs to be calibrated before the next calibration date.
§ If the next calibration date has passed, a dialog box appears telling the operator that the
transducer needs to be calibrated.
8.3.7 Product information
Re-open menu block “Conf.” (for programming tree, see 7.5 Configuration) and
select 5. Product information.
This is an information dialog box displaying the serial number, hardware revision,
software revision and calibration date on your ACTA 3000.
Information shown:
ACTA 3000 serial number
Hardware revision
Software revision
ACTA 3000 calibration date.
Next calibration date for the ACTA 3000.
Note!
When the ACTA 3000 is powered up, a check of the next calibration date is performed:
§ If there are less than 30 days to the next calibration date, a dialog box appears telling the
operator that the ACTA 3000 needs to be calibrated before the next calibration date.
§ If the next calibration date has passed, a dialog box appears telling the
operator that the ACTA 3000 needs to be calibrated.