46 User Manual Synertia® RF Generator
The information contained in this manual is subject to change by Comet Yxlon without prior notice.
4.5 Connecting RF power cable
INFORMATION
Make sure that high quality coaxial cables and connectors are used on all connections.
Safety note 30: Information - RF cables
WARNING
High voltage and radio frequency hazards!
Do not touch the RF Out connector when the generator is switched on.
Safety note 31: Warning - RF hazards
NOTICE
When mounting the connector to the back of the generator, mind the
following maximum torques:
7-16 panel jack:
• Coupling torque (recommended): 25 Nm to 30 Nm
• Proof torque: 35 Nm
N-type panel jack:
• Coupling test torque: max. 1.7 Nm
• Recommended torque: 0.7 Nm to 1.1 Nm
Safety note 32: Notice - connectors
Figure 25: RF Out connector with non-depressed switch plate (air- and water-cooled generator)
Spring-loaded
switch plate