6.2 Troubleshooting guide
This chapter contains basic troubleshooting information. Before contacting Customer
Support, follow the steps in section 6.2.1 "General troubleshooting".
WARNING
Defeating the interlock function can result in dangerous situations if the
generator is allowed to supply power into the customer's system with
inadequate safety precautions.
Safety note 41: Warning - interlock
6.2.1 General troubleshooting
Prior to specific troubleshooting procedures, verify the following:
Verify that you installed the Synerti RF generator according to the installation
instructions in this manual. See section 4.1 on page 39.
Verify that the AC Mains voltage is within specifications. See section 4.7 on
page 52.
6.2.2 Checks with the power off
DANGER
Risk of death or bodily injury!
Disconnect and lockout/tagout all sources of input power before working
on this unit or anything connected to it.
Safety note 42: Danger - high voltage
Ensure that the power to the unit is off.
Check for visible damage to the unit, cables, and connectors.
Ensure that all unit connectors are installed correctly and fastened tightly.
Check to determine whether any system-related circuit breaker has been tripped.
Verify that the input power to the unit meets specifications.
Ensure ground connections are adequate and secure.