EasyManua.ls Logo

Comet Synertia RF - Figure 28: Fitting with Teflon Sealing Tape

Comet Synertia RF
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual Synertia® RF Generator 49
The information contained in this manual is subject to change by Comet Yxlon without prior notice.
You need two fittings with NPT 3/8 male thread e. g. hose nozzle and
commercially available Teflon® sealing tape.
Wind two to maximum three layers of Teflon® sealing tape tightly onto the
threads.
Observe the winding direction (see
Figure 28).
NOTICE
The Teflon® tape should not overlap the start of the threads.
Safety note 35: Notice - Teflon tape
Figure 28: Fitting with Teflon® sealing tape
Hand-tighten the fittings (approx. three turns) on both water connections and
then tighten them with a spanner (two to three turns).
NOTICE
To avoid damages to the fittings, the maximum tightening torque
is 30 Nm.
Safety note 36: Notice - tightening torque