User Manual Synertia® RF Generator 55
The information contained in this manual is subject to change by Comet Yxlon without prior notice.
The external disconnect device must:
be rated for at least 10 kAIC as short-circuit breaking capacity,
be compliant to standards IEC 60947-1 and IEC 60947-3,
not be installed in the mains cable,
be readily accessible by the user,
be located such that an operator or inspector of the device is not
subjected to serious risk of tripping, falling, or being exposed to electrical
or mechanical hazards,
be lockable only in the "OFF" or isolated position,
be marked as the disconnect device for the equipment.
WARNING
To prevent potential life-threatening danger, a switch or circuit
breaker must not interrupt the protective conductor.
Safety note 39: Warning - protective conductor