F15 Series Fetal & Maternal Monitor User Manual Alarms
- 59 -
CTG Alarm(Low Priority)
FHR1/FHR2/FHR3
Suspected Delayed
Deceleration
Delayed deceleration is
detected when maternal and
fetal heart rate are close, and
the number of delayed
decelerations is >0 within 9 min
Check the patient’s and fetus’s current
condition.
Technical Alarm Messages
Alarm Message
Cause
Countermeasure
Medium Priority
**Battery Low
The battery indicator of the monitor
is . The battery power is too
low to support further work of the
monitor.
Connect the monitor to AC power
supply.
** Battery Failure
The battery is defective or
damaged, and the effective status
of the battery cannot be obtained.
Please contact the manufacturer for
service
**The battery level is
low(US1)
The battery indicator of wireless US
transducer 1 is .The battery
power is too low to support further
work of the transducer.
Charge the wireless US transducer
1 in the docking slot.
**The battery level is
low(US2)
The battery indicator of wireless US
transducer 2 is . The battery
power is too low to support further
work of the transducer.
Charge the wireless US transducer
2 in the docking slot.
**The battery level is
low(US3)
The battery indicator of wireless US
transducer 3 is . The battery
power is too low to support further
work of the transducer.
Charge the wireless US transducer
3 in the docking slot.
**The battery level is
low(TOCO)
The battery indicator of wireless
TOCO transducer is . The
battery power is too low to support
further work of the transducer.
Charge the TOCO transducer in the
docking slot.
**The battery level is
low(DECG)
The battery indicator of wireless
DECG Fetal&Maternal Module
is . The battery power is too
low to support further work of the
transducer.
Charge the DECG Fetal&Maternal
Module in the docking slot.
**The battery level is
low(FT20)
The battery indicator of FT20
Module is . The battery power
is too low to support further work of
the module.
**The battery level is
low(FECG)
The battery indicator of FECG
Fetal&Maternal Module is .
The battery power is too low to
support further work of the
Charge the FECG Fetal&Maternal
Module in the docking slot.
Traditional 510(k) of Fetal & Maternal Monitor
014_14.1_F15_Series_User_Manual
75202

Table of Contents

Other manuals for EDAN F15 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F15 Series and is the answer not in the manual?

EDAN F15 Series Specifications

General IconGeneral
BrandEDAN
ModelF15 Series
CategoryMonitor
LanguageEnglish

Related product manuals