F15 Series Fetal & Maternal Monitor User Manual Alarms
- 59 -
FHR1/FHR2/FHR3
Suspected Delayed
Deceleration
Delayed deceleration is
detected when maternal and
fetal heart rate are close, and
the number of delayed
decelerations is >0 within 9 min
Check the patient’s and fetus’s current
condition.
Technical Alarm Messages
The battery indicator of the monitor
is . The battery power is too
low to support further work of the
monitor.
Connect the monitor to AC power
supply.
The battery is defective or
damaged, and the effective status
of the battery cannot be obtained.
Please contact the manufacturer for
service
**The battery level is
low(US1)
The battery indicator of wireless US
transducer 1 is .The battery
power is too low to support further
work of the transducer.
Charge the wireless US transducer
1 in the docking slot.
**The battery level is
low(US2)
The battery indicator of wireless US
transducer 2 is . The battery
power is too low to support further
work of the transducer.
Charge the wireless US transducer
2 in the docking slot.
**The battery level is
low(US3)
The battery indicator of wireless US
transducer 3 is . The battery
power is too low to support further
work of the transducer.
Charge the wireless US transducer
3 in the docking slot.
**The battery level is
low(TOCO)
The battery indicator of wireless
TOCO transducer is . The
battery power is too low to support
further work of the transducer.
Charge the TOCO transducer in the
docking slot.
**The battery level is
low(DECG)
The battery indicator of wireless
DECG Fetal&Maternal Module
is . The battery power is too
low to support further work of the
transducer.
Charge the DECG Fetal&Maternal
Module in the docking slot.
**The battery level is
low(FT20)
The battery indicator of FT20
Module is . The battery power
is too low to support further work of
the module.
**The battery level is
low(FECG)
The battery indicator of FECG
Fetal&Maternal Module is .
The battery power is too low to
support further work of the
Charge the FECG Fetal&Maternal
Module in the docking slot.
Traditional 510(k) of Fetal & Maternal Monitor