STP-iXA4507 series Turbomolecular Pump
SERIAL COMMUNICATION PROTOCOL
1. In the case of time flag = 0
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
32-bits hexadecimal
coded ASCII
Control unit running time
32-bits hexadecimal
coded ASCII
*1 The values corresponding to the error message are transmitted (refer to Table 24, Table 25). When the number of
detected errors is under the maximum number, the parameter value larger than the number of detected errors is set to
FF
hex
.
Example:
In the case of Disturbance X_H detection at
Pump running time = 5,000 minutes, control unit running time = 6,000 minutes
2. In the case of time flag = 1
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
Error detection time
(Format:yymmddhhnn)
yy: The last two digits of the year
mm: Month
dd: Day
hh: Hour (24-hour display)
nn: Minute
40-bits hexadecimal
coded ASCII
24-bits hexadecimal
coded ASCII
*1 The values corresponding to the error are transmitted (refer to Table 24, Table 25). When the number of detected errors is
under the maximum number, the parameter value larger than the number of detected errors is set to FF
hex
.
*2 Each value of time is transmitted as BCD format character string.