STP-iXA4507 series Turbomolecular Pump
SERIAL COMMUNICATION PROTOCOL
Example:
In the case of Disturbance X_H detection on September 13, 2007, 12: 34
*3 System reservation
Example (in the case of time flag = 0):
When 3 errors have been detected in the past;
The number of "Error Record" :
The maximum number of "Error Record" :
Error Code: 0F
hex
= 15 = Disturbance Xb
Time flag: 00
hex
= Detection time is total running time
Pump running time = 5,000 hrs = 300,000 minutes
Control unit running time = 6,000 hrs = 360,000 minutes
Error Code: 0D
hex
= 13 = Disturbance Xh
Time flag: 00
hex
= Detection time is total running time
Pump running time = 700 hrs = 42,000 minutes
Control unit running time = 800 hrs = 48,000 minutes
Error Code: 12
hex
= 18 = MOTOR Overheat
Time flag: 00
hex
= Detection time is total running time
Pump running time = 20 hrs = 1,200 minutes
Control unit running time = 30 hrs = 1,800 minutes