GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-8 - Replacing Main Console 8 - 143
8-8-15 Replacement of the Rear Cover Top Assy-U
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-8-15-1 Preparations
Remove the Rear Cover Mid Assy-T as described in 8-8-10 "Replacement of the Rear Cover Mid Assy-
T" on page 8-138
8-8-15-2 Removal Procedure
8-8-15-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”
as described in Section 8-
12 on page 8-223
Table 8-77 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 5 minutes No tools required
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Pull the Rear Cover Top Assy-U to remove from the
console.