Microlab
®
STAR Service Manual
6-13
Adjustment Macro Programs (continued)
This List follows the order of Adjustment and Calibration
File Name Description
PIP ALIGNMENT WITH LLD.MCR With Channel Calibration Tool
173960 and Low Volume Tips.
Final visual Pipetting Channel check.
Checks alignment of all channels
with Low Volume Tips picked up.
Measurement Macro Program.
AUTOLOAD AUTOADJUSTMENT.MCR With Autoload Calibration Tool
173975 and Autoload Sensors
Adjustment Tool 173980.
Calibrates positioning in X, Y and Z
Direction automatically after
mechanical Cog wheel
synchronization.
Checks Barcode reading, Cup
Presence sensor adjustment and the
quality of Carrier loading and
unloading.
Calibration Macro Program.
Additionally for ML STAR IVD
Tightness check procedure With waste block 185281 and user
Software ML STAR IVD
Especially after replacement of
CORE O-Rings the tightness of seal
on each Pipetting Head may be
checked.
PIP_PRESSURE_SIGNALS_CHECK.M
CR
and
PIP_PRESSURE_SIGNALS_AUTOADJ
USTMENT.MCR
With waste block 185281 and
pressure Transmitter the values for
TADM may be checked and
adjusted.