149
Saving Data
Manual saving (selective saving, immediate saving)
This section describes how to save data using the SAVE key.
You can select the operation to perform when the SAVE key is pressed.
IMPORTANT
Data can be saved while the instrument does not perform measurement has. Data cannot be
saved while measurement is in progress.
Only data saved in the internal bu󰀨er memory (the last 256 Mwords) can be saved. Use auto
(real-time) saving if you need to save more than 256 Mwords of data.
> >
2
2
3
3
4
4
1
1
1
Under [SAVE key settings], select the operation to perform when the SAVE key is pressed.
Select & Save
Displays the settings window when the SAVE key is pressed.
Data will be saved after you select what to save.
Quick save Saves data immediately when the SAVE key is pressed.
The type of data set with [Type] will be saved.
2
Enter the lename to use when saving data in the [File name] eld (up to 8 single-byte
characters).
See “Text entry method” (p. 10).
If you do not specify a lename, a lename will be assigned automatically.
See “When saving les without specifying a lename” (p. 136).
For more information about le names and [Add trigger date/time], see [File name] of selective
save (p. 152).
3
Under [Media], select the media on which to save data.
SD card
, USB ash drive
Saving and Loading Data