Performing Remote Operation Using the HTTP Server
Remote version update
You can update the instrument and modules through a web browser.
When the instrument and modules need to be updated, Hioki will release the updating les and
version update procedure on the company’s website.
To update the instrument and modules remotely, download a compressed folder from the website
and extract it in advance.
Neither an SD Memory Card nor a USB Drive is required.
4
4
5
5
3
3
2
2
1
Launch Internet Explorer
®
on the computer.
2
Enter the instrument’s address into the address bar (for example, ).
3
Click [UPDATE].
4
Click [Browse...]. Select the update les you downloaded (The .UPG le and .PRG le is
used for the instrument and the modules, respectively).
5
Click [Update].
Executing the update of the instrument will automatically perform a restart, interrupting
communications.
For more information about cautions and an after-update checklist, see the procedure.
HIOKI LR8450A964-03
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Hioki LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki LR8450 and is the answer not in the manual?

Hioki LR8450 Specifications

General IconGeneral
ModelLR8450
CategoryData Loggers
Dimensions330 mm (12.99 in) W × 221 mm (8.70 in) H × 120 mm (4.72 in) D
WeightApprox. 2.8 kg (excluding battery)
Operating Temperature0 to 40 °C (32 to 104 °F), 80% RH or less (no condensation)
Sampling Rate1 ms to 1 hr
MemoryInternal memory
InterfaceUSB, Ethernet
Power SupplyAC adapter or battery
Measurement RangeVaries with module
AccuracyVaries with module
Storage MediaSD card
Memory CapacityInternal memory
Input TypeVoltage, temperature
Temperature MeasurementVaries with module
Communication InterfaceUSB, Ethernet
Measurement ChannelsUp to 120 channels (with optional input modules)