EasyManua.ls Logo

Hioki LR8450 - Auto-Setup Function

Hioki LR8450
425 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
159
Loading Data
Overwrite mode
Only when the conguration of the plug-in modules remains the same as that had been applied
when the data was saved, the le can be loaded.
Attempting to load a le with the di󰀨erent conguration of the plug-in modules will cause a loading
error.
View mode
When loading a le in View mode, the instrument will save the present settings in the internal bu󰀨er
memory.
The setting condition will temporarily change; however, exiting View mode will enable the previous
settings to restore.
The instrument cannot start measurement in View mode.
How to exit View mode
To exit View mode, execute any one of the following actions:
Press the ENTER key while [Exit VIEW mode] on the le list screen is selected.
Press the START key. (The instrument will start measurement after restoring the previous
setting.)
Press the MONITOR key.
Load the le in overwrite mode.
Turn o󰀨 the instrument. Initialize the instrument. Send the communications command.
Auto-setup function
This section describes how to automatically load the setting le on startup.
Save the [STARTUP.SET] le in the [CONFIG] folder (under [HIOKI] > [LR8450]). The instrument
will automatically load the le.
If both the SD Memory Card and the USB Drive contain the [STARTUP.SET] le, the instrument will
preferentially load the le on the SD Memory Card.
The instrument can automatically load the settings saved in the instrument’s internal backup
memory.
See “Saving settings to the instrument’s internal backup memory” (p. 155).
Saving and Loading Data
HIOKI LR8450A964-03
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents

Other manuals for Hioki LR8450

Related product manuals