EasyManua.ls Logo

Hioki LR8450 - Measuring Logic Signals

Hioki LR8450
425 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
43
Conguring Input Channels
Measuring logic signals
This section describes how to congure settings on the individual settings screen when measuring
logic signals.
You can use the settings list screen to congure the settings. (See p. 69.)
External control terminals: Pulse input terminals P1 to P8
> >
4
4
1
1
3
3
5
5
2
2
1
Select [Pulse], and choose a channel from [P1] to [P8],
Measurement will not be performed for channels whose check boxes are not selected.
2
Select the waveform display color.
× (OFF), 24 colors
Select [×] if you wish to measure the channel but not to display its waveform on the screen.
3
Set the input type to [Logic].
4
Under [Threshold], select the level to count.
1 V
Treats voltages that are greater than or equal to 1.0 V as high level and voltages that
are greater than or equal to 0 V but less than 0.5 V as low level.
4 V Treats voltages that are greater than or equal to 4.0 V as high level and voltages that
are greater than or equal to 0 V but less than 1.5 V as low level.
5
Under [Filter], select whether to use the chatter prevention lter.
Set to [ON] to prevent false counting due to chatter in mechanical contact (relay) output.
OFF
, ON
Settings and Operation
HIOKI LR8450A964-03
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents

Other manuals for Hioki LR8450

Related product manuals