121
Analog Triggers, Pulse Triggers, Waveform Calculation Triggers
Level triggers
Level triggers activate when the waveform crosses the specied level (the trigger level).
You can set the direction in which the level is crossed (the slope).
> > [Unit n], [Remote n] (n = 1, 2, . . .)
2
2
3
3
1
1
1
1
1
Under [Type] for [Start] or [Stop], set the trigger type to [Level].
A level trigger will be used.
2
Select the slope.
The trigger will activate when the level is crossed in the set direction.
Activates the trigger when the level is crossed the specied level from below. (Rising)
Activates the trigger will activate when the level is crossed the specied level from
above. (Falling)
When [Condition] is set to [AND], the system will determine whether the waveform has exceeded the
specied level.
3
Under [Level], set the trigger level.
The trigger will activate when the set level (temperature, voltage, etc.) is crossed.
When using the scaling function, post-scaling values are displayed.
(When a strain measurement module is used)
The instrument expresses strain in terms of micro epsilon (μ
ε
). You do not need to enter the SI
prex micro (μ).
Trigger Function
HIOKI LR8450A964-03
w ww . . co m
information@itm.com1.800.561.8187