217
Controlling the System
Synchronizing the time
This section describes how to synchronize the instrument’s time with an NTP server.
The LAN settings must be congured in advance.
See “9.3 Conguring and Establishing a LAN Connection” (p. 239).
> >
1
1
2
2
3
3
4
4
6
6
7
7
5
5
1
Press the ENTER key while [NTP settings...] is selected.
The settings window will open.
2
Set the [NTP client function] to [ON].
3
Under [Sync timing], select when to connect to the NTP server and set the time.
OFF
Does not set the time.
Every hour Sets the time once an hour.
Every day Sets the time once a day.
4
Under [Sync before start], select whether to connect to the NTP server and set the time
before starting measurement.
OFF
, ON
5
Under [Server address], set the address of the NTP server.
6
Press the ENTER key while [Sync now] is selected.
Once the settings have been applied, the instrument will connect to the specied NTP server and
set the time.
7
Press the ENTER key while [Apply] is selected.
The settings will be applied.
Press the ENTER key while [Cancel] is selected to close the window without changing any
settings.
Conguring System Settings
HIOKI LR8450A964-03
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Hioki LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki LR8450 and is the answer not in the manual?

Hioki LR8450 Specifications

General IconGeneral
ModelLR8450
CategoryData Loggers
Dimensions330 mm (12.99 in) W × 221 mm (8.70 in) H × 120 mm (4.72 in) D
WeightApprox. 2.8 kg (excluding battery)
Operating Temperature0 to 40 °C (32 to 104 °F), 80% RH or less (no condensation)
Sampling Rate1 ms to 1 hr
MemoryInternal memory
InterfaceUSB, Ethernet
Power SupplyAC adapter or battery
Measurement RangeVaries with module
AccuracyVaries with module
Storage MediaSD card
Memory CapacityInternal memory
Input TypeVoltage, temperature
Temperature MeasurementVaries with module
Communication InterfaceUSB, Ethernet
Measurement ChannelsUp to 120 channels (with optional input modules)

Related product manuals