217
Controlling the System
Synchronizing the time
This section describes how to synchronize the instrument’s time with an NTP server.
The LAN settings must be congured in advance.
See “9.3 Conguring and Establishing a LAN Connection” (p. 239).
> >
1
1
2
2
3
3
4
4
6
6
7
7
5
5
1
Press the ENTER key while [NTP settings...] is selected.
The settings window will open.
2
Set the [NTP client function] to [ON].
3
Under [Sync timing], select when to connect to the NTP server and set the time.
OFF
Does not set the time.
Every hour Sets the time once an hour.
Every day Sets the time once a day.
4
Under [Sync before start], select whether to connect to the NTP server and set the time
before starting measurement.
OFF
, ON
5
Under [Server address], set the address of the NTP server.
6
Press the ENTER key while [Sync now] is selected.
Once the settings have been applied, the instrument will connect to the specied NTP server and
set the time.
7
Press the ENTER key while [Apply] is selected.
The settings will be applied.
Press the ENTER key while [Cancel] is selected to close the window without changing any
settings.