EasyManua.ls Logo

Hioki LR8450 - External Trigger Input

Hioki LR8450
425 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
229
Conguring External Input/Output (I/O) Terminals
External trigger input
This section describes how to input a signal from an external source to activate a trigger.
You can use a signal from another device to start measurement with the instrument. The external
input I/O 3 terminal is used for this functionality.
> >
1
1
2
2
1
Under [External input 3], set the terminal functionality to [Trigger input].
When [External trigger] is set to [ON], the parameter will be set to [Trigger input].
See “2.6 Applying Triggers Based on External Sources” (p. 129).
2
Select the edge at which to activate the trigger.
Activates the trigger at the rising edge when changing from low level to high level.
Activates the trigger at the falling edge when changing from high level to low level.
The trigger will also activate if the I/O 3 and GND terminals are shorted.
It is recommended to actually input the signal and verify whether the external trigger operates.
External Control (EXT. I/O)
HIOKI LR8450A964-03
w ww . . co m
information@itm.com1.800.561.8187