6 - 2
(2) SW2
SW No. Function Description
1 Sample transfer ON : With sample transfer
2 External personal computer ON : With external personal computer
3 Unused
4 Sample ID ON : With sample ID
5 ISE ON : With ISE
6 Unused
7 Unused
8 Unused
6.4 LOG AMP (P/N 707-5009)
SW No. Type Setting Remarks
SW1 Rotary
O
Selection of main wavelength
SW2 Rotary
O
Selection of sub wavelength
SW3 Toggle
C
Multiplexer (F: manual setting)
6.5 ISE AMP (P/N 707-5023)
SW No. Type Setting Remarks
SW1 Rotary
O
Address setting
C F