2 - 63
(1) List of Calibration Output Media
Output Medium
Output Data
CRT
(screen name)
PRT S. I/F
Calibration curve
parameters
(S1ABS, K, A, B, C)
CALIBRATION LIST
× ×
SD value
×
CALIBRATION MONITOR
O
Serum index blank CALIBRATION LIST
× ×
Blank level
× × ×
Absorbance or absorbance
change rate
×
CALIBRATION MONITOR
O
Initial or final observance
×
CALIBRATION MONITOR
O
NOTE: In plotting of measured absorbance values, the currently measured STD alone is
taken.
(2) Relationship between Alarm and Output in Calibration
Data Alarm Name
Updating of Screen
Parameter
Screen Display
of Alarm
Remarks
S1ABS error No No
DUPLICATE error No No
STD error No Yes
SENSITIVE error No Yes
CALIB error Yes Yes
SD error Yes Yes
No alarm Yes No
NOTE: For CALIB and SD errors, "∗" is output on logging for discrimination.