7 - 7
7.1.6 Operation
(1) Operation (sample probe)
1. Function
• To discharge sample into reaction cell (photometry and ISE assays)
• To discharge rinse water into reaction cell
2. Sample probe status transition
NOTES: 1. When STD (1) POS#99 is specified in analytical parameters .
2. Sample pipettiing volume conpensation :
Actual pipetting volume "yµL" versus value input from screen "xµL" is calculated via following formula.
3. One addional pulse is applied for sample aspiration and discharge .
X(µ1) y(µ1)
20.0 to 15.5 x
15.6 to 25.5 1.04788 × -0.7109
25.6 to 50.0 1.03160 × -0.2479