16 - 40
(3) Test Selecting Information (from host to analyzer)
Send test selecting information corresponding to sample information.
ISE channel
Photometry-assay channel
Channel count
Details of Request for Each Channel
(b: Space)
Function Character Details of Request
Initial measurement " Ab " 0: No request
" b " 1: Normal sample volume
" Db " 2: Unused
3: Unused
4: Volume determined by analyzer side
(left to analyzer)
NOTES: 1. The above channel count ("bb0" to "b37") is the number of effective
channels from photometry-assay channel 1. If "bb3" is specified,
channels 1 to 3 are reflected on the analyzer with channels 4 to 37
ignored.
When there is at least one requested test, it is desirable to set "37."
Example:
Not reflected on analyzer (analyzer side is followed)
Reflected on analyzer
Channel count
The analyzer side is obeyed if test request is made in a channel number
beyond the specified channel count. To avoid analysis for an
unnecessary test, set channel count "37" and request for only a test to be
specified.
2. Request for ISE is specifiable in the 37th channel.
(i) Specification of any other than "0" ... Select Na, K and Cl on the
screen.
(ii) Specification of "0" ... No request
Note that it is impossible to select a request for any of Na, K and Cl from
the host.
1 2 3 4 5 ... 36 37
Transfer sequence
1 2 3
4
5 36 37
b b 3 1 1 1 1 1 .… 0 0 1