12 - 20
(6) Memory Initialization
[Memory Initialization Menu]
[Menu End]
[Image Data Initialization]
Select [Image Data Initialization].
Is it desired to initialize image data
memory?
[Yes]
[No]
Select [Yes].
The image data memory is initialized.
"Initialization is under execution." is displayed during initialization.
Upon completion of initializing the image data memory, execute installation according to
"Procedure for Installation of Product Program".
(7) Memory Switch Setting 1
(NOTE: Each switch is set as shown below. Setting change may disable startup of the
instrument. So avoid changing the setting.)
[Keying Sound] OFF Stop
[Buzzer Sound] ON Next
[Back Light Off Time] 10 Min Screen
[Automatic Commu- Write
nication Recovery] No Check
Press [Write Check] key.
(8) Memory Switch Setting 2
(NOTE: Each switch is set as shown below. Setting change may disable startup of the
instrument. So avoid changing the setting.)
[Data Bit Length] 8 Bits Stop
[Stop Bit Length] 2 Bits Previous
[Parity] Even Screen
[Flow Control] RS/CS
[Baud Rate] 9600 bps Write
[Lamp Bit Control] Invalid Check
Press [Write Check] key.