7 - 46
(1) Precision Check
(a) Statitical calculation is made for the result data of routine samples.
(b) Specify Precision Check on the MAINTENANCE screen.
(c) Statistical data for a maximum number of samples is calculated. Up to forty tests
including ISE assay are applicable.
(d) The statistical data shown in the following table is output.
Output Statistical Data Output Format
Number of Digits below
Decimal Point
Data Count (N) 3-digit integral −
Maximum value (MAX) 6-digit real number with
sign/decimal point
According to STD(1) CONC
Minimum value (MIN) Same as above Same as above
Mean value (MEAN) 7-digit real number with
sign/decimal poin
1 digit increased below
decimal point as compared
with STD(1) CONC
Standard deviation
(SD)
Same as above Same as above
Coefficient of variation
(CV)
6-digit real number with decimal
point
Fixed at 2 digits
(e) The result data is read from FD.
(f) The result data provided with an alarm is excluded from calculation (also not
counted for N). Note that reference value range over and overflow are excluded
from alarm.
(g) Statistics are not output for unmeasured tests and subsequent tests are placed
close.
(h) When N < 2, statistical calculation is not carried out.