2 - 65
2.5.1 Logging Program List
Printout Stop
by Stop Key
1 Monitor printout Real time printout: Specify "print" on
the start condition menu screen.
Printout is made when the specified
values in all tests for one sample are
calculated during operation.
Batch printout: Specify sample
number on the data monitor screen.
Real time
and batch
O
(In batch mode
alone)
2 Calibration result
printout
Printout is made when "calibration" is
selected on the start condition menu
screen and the result of calibration is
output.
Real time
×
3 Remaining
reagent volume
printout
Specify "remaining reagent volume
printout" on the start condition menu
screen.
Batch
×
4 Cell blank printout Specify "cell blank" on the
maintenance screen.
Batch
O
5 Reproducibility
check printout
Specify "reproducibility check" on the
maintenance screen.
Batch
O
6 Cumulative
mechanism
information
printout
Specify "cumulative mechanism
information" on the mechanism check
menu screen.
Batch
×
7 Parameter printout Specify "parameter" on the
maintenance screen.
Batch
O
8 Program check
printout
Specify "program check" on the
mechanism check menu screen.
Batch
×
9 ISE check printout Specify "ISE" check on the mechanism
check menu screen.
Real time
×
10 Photometer check
printout
Specify "photometer check" on the
maintenance screen.
Batch
O
(Note 1)
11 Alarm trace
printout (DAILY)
Specify "alarm trace information" on
the mechanism check menu screen.
Batch
O
12 Alarm trace
printout
(CUMULATIVE)
Specify "alarm trace information" on
the mechanism check menu screen.
Batch
O
13 Original Abs.
printout
Printout is made when "original Abs." is
specified on the system parameter
screen and operation is executed.
Real time
×
(Note 2)
14 Communication
trace printout
Specify "communication trace printout"
on the mechanism check menu screen.
Batch
O
15 Work sheet
printout
Specify "work sheet printout" on the TS
registration screen of routine analysis.
Batch
O
16 Barcode Reader
check
Specify "Barcode Reader" check on
the mechanism check menu screen
Real time
×
NOTES: 1. The stop key is invalid during printout.
2. The stop key need be pressed to stop analysis.
(1) Initialization of Printer