7 - 45
7.5.7 Photometer Check
Function : Absorbances at 12 wavelengths are measured and output onto the p rinter
together with the previously measured data.
Required time: 1 min 50 sec (reset 20 sec + 18 sec/cycle × 5 cycles)
Processing
No.
Details of Processing Reference Document
1) Start from MAINTENANCE screen Screen specifications
2) Refer to the next page.
3) Absorbances at 12 wavelengths (main/sub) are printed
out together with previous ones stored in CRAM.
The recent result data is stored in CRAM.
Photometer report data check
4) At execution of photometer check, operation is checked
and alarm is displayed if operation is abnormal.
Screen specifications
Alarm code specifications
5) The above alarm is stored in FD. FD specifications
Remarks
Judgement
by operator
START
MAINTENANCE
screen
Photometer
check required
Photometer data
print
Storage of
photometer data
Memory
Alarm indication
Data output
Operation check
Abnormal
Normal
FD
(alarm logging)
END
Input of "Execute"
No
Yes
1)
2)
3)
5)
4)
Photometer check