2 - 18
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
4811 to
4848
VOLUME
CHECK?
100 1 to 36
(ch.)
WARNING (1)The total reagent
volume up to the last
photometric point is
more than 250 to
500 µL.
(2)The volumes of
reagents 1 to 3 are all
zero.
(3)The reagent volume
having a timing behind
the reaction time is not
zero.
(4)The total liquid volume
of sample and reagent is
less than 250 µL.
1.When '999' (stirring
only) is specified for
reagent volume, 0 µL is
taken for the volume.
However, when '999' is
specified for the reagent
volume having a timing
behind the reaction time,
an error occurs.
2.Operation is
unallowable.
3.The final liquid volume
is a total volume of
sample and reagent
within 250 to 500 µL.
Check parameter.
If the improper
condition indicated
by alarm can be
detected, correction
and reentry are
required.
5261 CMP. TEST 103 1 to 8 WARNING The setting of formula
number corresponding to
the relevant code is
improper.
(1)An unmeasurable test is
specified for
compensation.
(2)A compensated test is
not included in the
formula.
(3)In photometry assay for
compensated test, the
electrolyte parameter is
specified.
Call up calculation
item screen and
check the
compensation
formula on it.
(cont’d)