2 - 17
(cont’d)
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
4511 to
4548
CHEMISTRY
PARAME-
TER?
98 1 to 36
(ch.)
WARNING(1)The relationship
between assay code and
photometric point is
improper.
(2)The assigned
photometric point lags
behind the specified
reaction time.
(Operation is
impossible.)
Correct the
parameter.
4661 to
4698
CALIBRA-
TION
PARAME-
TER CALIB.
99 1 to 36
(ch.)
WARNING (1)The relationship
between assay code and
calibration type is
improper.
(2)Necessary calibration
points for calibration
type are not input.
(3)Necessary standard
positions for calibration
are not input.
(4)The relationship
between calibration type
and calibration method
is improper.
Check parameter
and input it again.
(5)The standard
concentration values are
not set in ascending
order. (Except for STD
(3) and (4) for isozyme)
(6)Concentration value is
not zero when '99' is
entered for POS. of STD
(1).
1.The photometry assay
data in other than
manual mode is
checked.
2.Operation is
unallowable.
3.Check in (4) is made
with regard to
specifications of time-
out calibration, lot-to-lot
calibration, bottle-to-
bottle calibration and
test selecting
information.