2 - 16
(cont’d)
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
4201 to
4203
ISE
REAGENT
SHORT
95 1 Unassigned
(Liquid level detection for
IS/DIL solution)
Replace the
reference electrode
solution with new
one.
Check the ISE
reagent volume.
95 2 Unassigned (Same as
above)
95 3 WARNING The volume of reference
electrode solution is 30 mL
or less.
4211 TWLN
TEST?
96 1 to 36
(ch.)
WARNING (1) when analyzing two
tests at a time, the
assigned method for
the corresponding test
is inadequate.
(2) Assignment of the
corresponding test is
being done or not
being done when
analysis for two tests is
not being made at the
same time.
(3) When analyzing two
tests at the same time,
the analytical
parameters are not
identical.
Sub-codes (1 to 36)
signify the channel
numbers.
Start setting in
order from the
tests that are
designated by the
Measured Point of
the first half of the
1 channel 2 Test
Analysis Method (3
Point, 1 Point and
Rate, Rate B)
When not
analyzing two tests
at the same time,
do not designate
'Two Test Analysis'
for the Analysis
Parameter.
With the Analysis
Parameter screen
for the applicable
test, unify all the
parameters below
for the
Simultaneous 2
Test Analysis Test.
Analytical Method
Reaction Time
Sample Volume
Reagent
Pipetting volume
(R1-R3)
Calibration
Method
Calibration Point
Standard solution
volume
Standard solution
position