2 - 15
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
4171 to
4173
ISE PREP. 92 1 WARNING Upon calibration, the slope
value is within the following
range.
Na : 45.0mV Slope value
49.9mV or
68.1mV Slope value
92 2 WARNING K : 45.0mV Slope value
49.9mV or
68.1mV Slope value
92 3 WARNING Cl : -39.9mV Slope value
-35.0mV or
-68.1mV Slope value
4181 to
4183
ISE SLOPE 93 1 WARNING (1) In the result of
calibration, the slope
value is within the
following range.
(2) The response
characteristic of
electrode is poor (in case
carry-over rate (A) is as
indicated below).
Na : (1)SLOPE < 45.0mV
(2)0.232 < A
Refer to alarm
codes 92-1 to 3.
93 2 WARNING K : (1)SLOPE < 45.0mV
(2)0.160 < A
Same as above
93 3 WARNING Cl : (1)SLOPE < -35.0mV
(2)0.490 < A
Same as above
4191 to
4193
ISE I.STD 94 1 WARNING The concentration of internal
standard solution (C(IS)) is
within the following range.
Na : C(IS) < 120.0mEq/L or
160.0mEq/L < C(IS)
(1) Replace STD
and carry out
calibration
again.
(2) Replace the
internal
standard
solution.
94 2 WARNING K : C(IS) < 3.0mEq/L or
7.0mEq/L < C(IS)
Same as above
94 3 WARNING Cl : C(IS) < 80.0mEq/L or
120.0mEq/L < C(IS)
Same as above
(cont’d)
Make sure that the
standard solution
and reagent are set
properly.
Make sure that the
standard solution is
free from
concentration or
deterioration.
Make sure that the
electrodes (Na, K,
Cl) are within their
guaranteed life.