2 - 30
2.3.2 Details of Parameter Check
Alarm
Display Subdivision
1 Check of
parameters
for twin test
simultaneous
analysis
The following are checked
for measurable (level 1,
Note 1) photometry tests.
(1) In case of twin test
simultaneous analysis,
the opposite test must
be specified properly.
Two tests must form
a unique pair, and
one of them alone
must designate the
other.
(2) In case of other than
twin test simultaneous
analysis, tests must not
be paired.
(3) In case of twin test
simultaneous analysis,
two tests must be
identical in all of the
following parameters.
Data mode
STD POS.
STD S. VOL
Assay code
Reaction time
Sample volume
Reagent volume
Calibration type
Calibration point
Span point
TWIN TEST? 1 to 37
(CH No.)
Details of this check
are given in "Details
of twin test
simultaneous
analysis" in 2.3.3.
In case of twin test
simultaneous
analysis, an alarm
is issued if one of
two tests is not
registered or the
manual mode is
selected.
NOTE:
1. The term
'measurable'
indicates either
of the following
two levels.
Level 1:
Test registered
and manual
mode not
selected
Level 2:
Same as above
and reagent
provided
2 Assay code
check
The following are checked
for measurable (level 1)
photometry tests.
(1) Relationship between
assay code and
photometric point
(check of photometric
point input and input
range)
Refer to the analytical
method table in 1.1.1.
CHEM
PARAM?
1 to 37
(CH No.)
Remarks
Designation
of Check
No.
Details of Check