9 - 12
9.3.2 Cell Blank Compensation
<Supplementary explanation>
The cell blank calculation method (Note) and compensated points vary with analytical method.
For details, refer to the classification of analytical methods.
NOTE: If an alarm is indicated for any of cell blanks 1 to 4, the following table will be used.
Number of
Alarms
Cell Blank Value
Data Alarm
(cell blank abnormal)
1 Mean value of 2 data without alarm Not indicated
2 Indicated
3 Indicated
Raw absorbance
at main
wavelength
Raw absorbance
at sub wavelength
Cell blank
compensation
Cell blank
compensation
Calculation of
cell blank value
Calculation of
cell blank value