1 - 2
1.2 Sampling System
Sample container : Hitachi standard sample cup
Sample tube ;diameter 13 to 17 mm
length 75 to 100 mm
Sample disk : 60 positions
Outer row;35 positions (routine and stat samples)
(with free adapter, barcode reader attachable)
Inner row;25 positions (standard + control samples (22),
rinse solutions (3))
Detergents : 3 kinds
(W1;for sample probe rinsing) HITERGENT
(W2;for sample carryover evasion) ISE RINSE SOLUTION
(W3;for sample carryover evasion) HICARRYNON
Sampling mechanism : Pipetter driven by stepping motor
2 to 50 µL of sample/test (in 0.1 µL steps)
Pipetting mode : Normal ; prescribed volume discharged at cell
bottom
Aqueous sample; system water solution for sample probe
internal rinsing usable as STD-1
Sample pre-dilution not performed
Insufficient sample detection: Resistance detection method
Sample ID : Barcode reader (option)
Following code shemes are usable in combination; CODE 39,
CODE 128, INTERLEAVED 2 OF 5, NW7 (Modulas 10,
Modulas 16)
Automatic rerun : Auto rerun function not provided
Stat sample analysis : Routine sample analysis interruptible for stat sample analysis
Sample carryover : Sample carryover evasion function available; 2 kinds of
detergents settable on sample disk inner row
Sample conveyance : Can be sampled directly from belt-line