EasyManua.ls Logo

Hitachi 902 - Sampling System

Hitachi 902
427 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 - 2
1.2 Sampling System
Sample container : Hitachi standard sample cup
Sample tube ;diameter 13 to 17 mm
length 75 to 100 mm
Sample disk : 60 positions
Outer row;35 positions (routine and stat samples)
(with free adapter, barcode reader attachable)
Inner row;25 positions (standard + control samples (22),
rinse solutions (3))
Detergents : 3 kinds
(W1;for sample probe rinsing) HITERGENT
(W2;for sample carryover evasion) ISE RINSE SOLUTION
(W3;for sample carryover evasion) HICARRYNON
Sampling mechanism : Pipetter driven by stepping motor
2 to 50 µL of sample/test (in 0.1 µL steps)
Pipetting mode : Normal ; prescribed volume discharged at cell
bottom
Aqueous sample; system water solution for sample probe
internal rinsing usable as STD-1
Sample pre-dilution not performed
Insufficient sample detection: Resistance detection method
Sample ID : Barcode reader (option)
Following code shemes are usable in combination; CODE 39,
CODE 128, INTERLEAVED 2 OF 5, NW7 (Modulas 10,
Modulas 16)
Automatic rerun : Auto rerun function not provided
Stat sample analysis : Routine sample analysis interruptible for stat sample analysis
Sample carryover : Sample carryover evasion function available; 2 kinds of
detergents settable on sample disk inner row
Sample conveyance : Can be sampled directly from belt-line

Table of Contents