EasyManua.ls Logo

Hitachi 902 - Barcode Reader Switch Setting; Settings of each SW at Shipment

Hitachi 902
427 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14 - 4
(3) Adjust VR4 on EMIO100 board so reaction bath water temperature is 37 ± 0.1°C. (Since
the heater is turned ON/OFF at around 6.0 V, the tester output should be about 6.0 V
when reaction bath water temperature is 37°C.)
(4) Adjust the reaction bath temperature indication with VR3.
NOTE: A YSI thermometer should be used to measure the reaction bath water
temperature.
NOTE: The measuring position for the YSI thermometer is midway between the sample
discharge position and reagent discharge position.
14.1.5 Barcode Reader Switch Setting
Set and/or check the DIP switches of the barcode reader in the following procedure.
In case of sample barcode reader: Use the settings made by the manufacturer.
Details of setting are given in Table 14-3.
Fig. 14-2 Barcode Reader DIP Switch Settings
< Settings of each SW at shipment >
SW1
SW2
EMOT200 board
1
2
3
4
5
6
7
8
1 2 3 4 5 6 7 8
ON
ON
SW1 SW2

Table of Contents

Related product manuals