16 - 41
3. In request for the isozyme test or compensation test, the other test
necessary for isozyme calculation or test-to-test calculation is
automatically supplied for analysis, and tests added for serum indexes are
transferred as well.
4. When request for the calculation test is made, judge and request the
channel for the test necessary for the calculation. When request for the
A/G ratio is made for example, send test selecting information considering
the channels for TP (total protein) and ALB (albumin).
5. TS request for serum indexes cannot be made for each sample.
For request, specify serum index test on the parameter screen and serum
indexes on the start condition screen.
(4) Comment Information (from host to analyzer)
Presence/absence of comment 5
Presence/absence of comment 4
Presence/absence of comment 3
Presence/absence of comment 2
Presence/absence of comment 1
Set "0" for all comment presence/absence information items.
Transfer sequence