1 - 3
1.3 Reagent System
Reagent disk : One disk, 40 positions
(20 positions each on inner and outer rows)
Reagent cooling : Cooling water circulating system (3 to 15°C) (option for China-
destined instrument)
Reagent bottle : 50 mL, 20 mL (adapter necessary)
(without barcode, concentrated reagent unusable)
Detergents : HITERGENT (50 mL); position 40 (fixed)
(for addition to reaction bath)
Detergent 1; for reagent probe, stirring rod rinsing, reagent
probe carryover evasion, reaction cell carryover
evasion
Detergent 2; for reagent probe carryover evasion, reaction cell
carryover evasion
(settable on reagent disk inner row)
ISE reagent : Internal standard solution(position 37 (fixed))
Diluent (position 38 (fixed))
Pipetting mechanism : Pipetter driven by stepping motor; 20 to 350 µL/test (in 1 µL
steps)
Reagent addition timing: 3 steps (0, 1.5 and 5 minutes); throughput is maintained even
when adding 3 reagents.
Carryover evasion : Rinsing for carryover evasion possible; throughput may decrease
by specifying carryover evasion.