12 - 3
12. INSTALLATION
12.1 902 Layout and Installation Conditions
Fig. 12-1 Instrument Layout
Installation Conditions
(1) Space : Instrument size 720 (W) × 720 (D) × 1085 (H)
(2) Weight : Approx. 190 kg
(3) Power requirements : 100 V AC, 50/60 Hz, 1.5 kVA
(4) Power receptacle : 3P grounded receptacle
If the receptacle is not grounded, there must be a grounding
terminal near it.
(5) Ambient temperature : 18 to 30°C (within ±2°C variation during measurement)
(6) Ambient humidity : 20 to 80% RH (non-condensing)
(7) Deionized water
supply unit : Water pressure 50 - 340 Kpa; 0.5 to 3.5 kgf/cm
2
(8) Water drain port : 50 mm or more in opening diameter, 100 mm or less above floor
(9) Heat dissipation : 1.3 kcal max.
* Deionized water supply unit
3P receptacle
Piping, 5 m long or less
Water drain port 3P grounded receptacle
500 or more
Power cable,
5 m long
720
1000 or more
300 or more100 or more
720
* Available optionally