EasyManua.ls Logo

Hitachi 902 - Page 43

Hitachi 902
427 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 - 31
(contd)
Alarm
Display Subdivision
2 Assay code
check
(2) Photometric points after
the specified reaction
time must not be set.
Refer to Note 6 in
"analytical method
table" of 1.1.1.
CHEM
PARAM?
1 to 37
(CH No.)
3 Calibration
parameter
check
The following are checked
for measurable (level 1)
photometry tests.
(1) Relationship between
assay code and
calibration type
Refer to "relationship
between calibration type
and analytical method"
in 2.1.
(2) Entry must be made for
CALIB. POINTS
necessary for calibration
type.
Refer to "output by each
calibration method and
check table" in 2.7.
(3) Entry must be made for
STD POS. necessary
for calibration.
For STD specified by
measured STD No. in
"output by each
calibration method and
check table" of 2.7, it is
checked whether STD
POS. is input or not.
CLB. PARAM? 1 to 37
(CH No.)
NOTES:
1. Checked
against the
specification of
calibration test
selecting
information.
Remarks
Designation
of Check
No.
Details of Check

Table of Contents

Related product manuals