2 - 32
(contd)
Alarm
Display Subdivision
3 Calibration
parameter
check
(4) The input value for
SPAN POINT must be
normal.
When calibration
type is LINEAR (2 to
6-point) or LOGIT-
LOG (3P or 4P)
1) The above input
value must not
equal 0.
2) The above input
value must not
exceed that for
CALIB. POINTS.
(5) Relationship between
calibration type and
calibration method
Refer to "output by each
calibration method and
check table" in 2.7.
(Note 2)
(6) STD CONC. must be
set in ascending order.
Check is made only for
necessary STD POS.
Note that isozyme STD
(3) and (4) are
excluded.
"=" is not regarded as in
ascending order.
(7) When '99' is entered for
POS. of STD (1),
CONC. must be 0.
CLB. PARAM? 1 to 37
(CH No.)
NOTES:
2. Checked
against the
specification of
calibration test
selecting
information.
4 Volume
check
The following are checked
for measurable photometry
tests.
(1) The total liquid volume
up to the final
photometric point must
be 500 µL or less.
(2) At least one of R1 to R3
volumes must not be 0.
(3) The reagent volume
having a timing behind
the reaction time is not 0.
(4) The minimum liquid
volume must be at least
250 µL.
VOLUME ? 1 to 37
(CH No.)
When reagent
volume = 0, the
relevant reagent is
not used.
Remarks
Designation
of Check
No.
Details of Check